SCANPEDIA

Glossary

SCANpedia - welcome to the SCANLAB GmbH encyclopedia.

Here you'll find explanations and definitions of key technical terms from the fields of optics, scanner technology, laser technology and laser processing.

ALL  | A | C | D | E | F | G | I | L | M | N | O | P | R | S | T | W

D

Drift

Drift is alteration of the image field over a long time period (from five minutes up to several hours).

At long time scales, mechanical and electronic effects occur that influence the image field. The sum of these effects is called drift. Because a large portion of these deviations is attributable to ambient temperature changes, those are evaluated separately. Drift values specified by SCANLAB are always valid for each scan axis individually.

The following specifications apply:

  • Long-term drift: The temperature-independent drift component is specified for different time regimes:
    - 8 hours, after 30-minute warm-up
    - 24 hours, after 3-hour warm-up
    Prerequisites here are constant ambient temperature and steady work load of the scan system.

  • Temperature drift: Ambient temperature fluctuations are specified in accordance with the magnitude of temperature change.

Temperature and long-term drift specifications consist of error values for offset and gain:

  • Offset: Shifting of the image field specified as an angle per temperature change (offset error).
  • Gain: Deviation in image field scaling specified as percent deviation per temperature change (gain error).

Systems with low drift

For drift-sensitive applications, systems with digital encoders (also see digital encoder technology) can be advantageous.
Be careful to maintain constant environmental conditions.